Technical parameters |
||||||
|
Model |
101-0A/101-0AB |
101-1A/101-1AB |
101-2A/101-2AB |
101-3A/101-3AB |
||
|
Application |
used for treatment of samples with high humidity and density |
|||||
|
Air circulation type |
side air flue and forced convection |
|||||
|
Performance |
temp. range |
RT+10~ 250℃ |
||||
|
temp. resolution |
0.1℃ |
|||||
|
temp. fluctuation |
±1℃ |
|||||
|
temp. uniformity |
±2.5% |
|||||
|
Structure |
chamber material |
A refers to cold rolled steel; AB refers to stainless steel |
||||
|
shell material |
spraying cold rolled steel |
|||||
|
thermal insulation material |
aluminum silicate fiber |
|||||
|
heating device |
stainless steel heating tube |
nichrome heating wire |
||||
|
rated power |
1.2kw |
1.6kw |
2.0kw |
4kw |
||
|
vent |
inner diameter 28mm*2(back), testing hole (top) |
|||||
|
Controller |
Temp. control |
double rows of digital tubes, PID |
||||
|
Temp. setting |
press the four keys slightly |
|||||
|
Temp. display |
actual temp.:displayed by 4 digit digital tube(row 1); |
|||||
|
timer |
0~9999minutes(with timing function) |
|||||
|
operation function |
running with data set and time set, and automatic shutdown based on setting |
|||||
|
program mode |
optional |
|||||
|
sensor |
Pt100 |
|||||
|
additional function |
deviation calibration, menu buttons lock, power outage compensation, and power outage memory. |
|||||
|
Safety device |
over-temperature alarm |
|||||
|
Specification |
working size |
350*350*350 |
450*350*450 |
550*450*550 |
600*500*750 |
|
|
exterior size |
652*472*587 |
752*472*687 |
852*572*786 |
862*622*985 |
||
|
packing size |
800*640*670 |
910*640*810 |
1010*740*870 |
1160*840*1190 |
||
|
working capacity |
43L |
71L |
136L |
225L |
||
|
shelf load bearing |
15kg |
|
||||
|
shelf bracket groove |
9layers |
13layes |
17layers |
21layers |
||
|
height between shelves |
25mm |
|||||
|
Power source( 50/60Hz)/Rated current |
AC220V/5.5A |
AC220V/7.3A |
AC220V/9.1A |
AC220V/18.2A |
||
|
NW/GW |
33/37kg |
45/50kg |
62/68kg |
77/96kg |
||
|
Accessory |
shelf |
2pcs |
||||
|
shelf bracket |
4pcs |
|||||
|
Optional device |
shelf, RS485 port, printer, recorder, external communication, remote control, program temperature controller, wireless message alert, U disk storage. |
|||||